Recall Z-0308-2007 ยท MKJ Automated External Defibrillators (Non-Wearable)

Philips Medical Systems

HeartStart FR2+ Defibrillator

DateDec 27, 2006 (initiated Dec 19, 2006)
StatusTerminated
ReasonPotential for shock treatment to be delayed or prevented due to board contamination that causes intermittent switch operation
Root causeProcess design
ActionOn 12/19/06 the firm sent letters via return receipt to all domestic customers. The letter advised customers of the issue and if the user encounters the problem, to press the button again with additional force. The users are to contact Philips for replacement. The devices can be used until the user receives a replacement.
Quantity2000 units
DistributionDistribution is worldwide.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K003565HEARTSTREAM FR2 AED, MODEL M3860A,M3861A
Agilent Technologies, Inc., Dec 20, 2000