ClearanceBrief
Recall Z-0215-2007 ยท MWI Monitor, Physiological, Patient (Without Arrhythmia Detection Or Alarms)

Welch Allyn Protocol, Inc

Propaq LT vital signs monitor, Model 802LTON

DateNov 28, 2006 (initiated Oct 22, 2006)
StatusTerminated
ReasonPotential to burn patient's finger. A Nellcor pulse oximetry (Sp02) sensor became hot when used with a Propaq LT.
Root causeOther
ActionA letter dated October 22, 2006 was sent to customers between October 25-27,2006. The letter advised of the problem and requested the customer contact the firm to arrange for return of the monitor to be serviced. The customer will be provided with a temporary replacement. The dealers were requested to notify their customers.
Quantitytotal of both models 1031
DistributionDevices were distributed to hospitals/medical centers and dealers worldwide.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K033378VITAL SIGNS MONITOR, PROPAQ LT
Welch Allyn Protocol, Inc., Mar 1, 2004