Recall Z-0299-2007 ยท LGW Stimulator, Spinal-Cord, Totally Implanted For Pain Relief

Advanced Bionics Corporation

Advanced Bionics Precision Linear Leads, part of the Precision Spinal Cord Stimulation System. Model Numbers: SC-2138-30, SC-2138-50, SC-2138-50T and SC-2138-70.

DateJan 10, 2007 (initiated Sep 21, 2006)
StatusTerminated
ReasonA small number of unimplanted Precision Linear Leads may have been assembled with incorrect electrode material. At high stimulation levels, the metal may corrode and dissolved metals may enter the patient.
Root causeOther
ActionIn United States, the clinician recall notification letters were distributed by Fed-Ex or U.S. Postal Service certified mail on 09/21/2006. For the other regions, the in-country representatives started providing notification to the clinicians on 09/21/2006. The recall notification letters will instruct physicians to immediately return the unimplanted units.
Quantity18
DistributionWorldwide.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.