Recall Z-0298-2007 ยท DXY Implantable Pacemaker Pulse-Generator

St Jude Medical CRMD

St. Jude Medical Identity SR Model 5172, Identity DR Model 5370, & Identity XL DR Model 5376; Pulse Generators (pacemakers)

DateDec 19, 2006 (initiated Oct 6, 2006)
StatusTerminated
ReasonSt. Jude Medical has identified a low-frequency anomaly in the software used in the APS III Model 3500/3510 and Merlin PCS Model 3650 programmers that can lead to incorrect reporting of battery voltage, expected battery longevity and Elective Replacement Indicator (ERI) status.
Root causeOther
ActionThe notification material was sent to physicians by FedEx. The advisory notification was sent by FedEx on Oct. 12, 2006, to arrive at the physicians' offices on Oct. 13th. In the US, acknowledgement of receipt of the notification material will be monitored by St. Jude Medical CRMD in Sylmar, California. For the rest of the world it will be by various St. Jude offices and local distributors with this information being fed back to St. Jude Medical CRMD in Sylmar. In addition, St. Jude Medical issued a press release on October 13, 2006. Plus information concerning this advisory can be found on
Quantity199,065
DistributionWorldwide in USA, ARGENTINA, AUSTRALIA, AUSTRIA, BARBADOS, BELGIUM, BERMUDA, BOLIVIA, BRAZIL, CANADA, CAYMAN ISLANDS, CHILE, CHINA, COLOMBIA, CYPRUS, CZECH REPUBLIC, DENMARK, ESTONIA, EUROPE, FINLAND, FRANCE, GERMANY, GREAT BRITAIN, GUATEMALA, HONG KONG, ICELAND, INDIA, ISRAEL, ITALY, JAPAN, JORDAN,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.