ClearanceBrief
Recall Z-0258-2007 ยท MWI Monitor, Physiological, Patient (Without Arrhythmia Detection Or Alarms)

Medrad Inc

Invasive Blood Pressure Interface Cables: a) Medrad 9500 Multigas monitor for MRI, Invasive Blood Pressure Interface Cable (Part # 3006314) for use with Abbott/Hospira IBP Transducer b) Invasive Blood Pressure Interface Cable (Part # 3008772) for use with Braun IBP Transducers. The product is an accessory to patient physiological monitor.

DateDec 15, 2006 (initiated Sep 29, 2006)
StatusTerminated
ReasonInterference problems: certain invasive blood pressure transducers connected to these cables are susceptible to interference if exposed to static magnetic field strengths exceeding 300 gauss. In certain orientations, the transducer causes the invasive blood pressure readings on the physiological monitor to display inaccurate readings.
Root causeOther
ActionThe recalling firm contacted all affected customers via telephone or letter dated November 20, 2006 to inform them of the problem and instruct them to return the cables or apply corrective labeling.
Quantity80 units
DistributionWorldwide distribution ---- including USA state of CA, CO, HI, IL, IN, MA, MD, NJ, NY, OH, PA, TX, and WA and countries of The Netherlands, Denmark, Australia, Brazil, Italy, and Japan.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K983250INVASIVE BLOOD PRESSURE (IBP) FUNCTION, MODEL 9500 MULTIGAS MONITOR
Magnetic Resonance Equipment Corp., Dec 12, 1999