Recall Z-0304-2007 ยท IYE Accelerator, Linear, Medical

Varian Medical Systems Inc

4D Integrated Treatment Console and Varis 1.4g Medical Charged Particle Radiation Therapy System; All 4D Integrated treatment console versions 8.0.15 and below (excluding v7.0.31) and all Varis Non-Varian Treatment versions 6.6.5027 and below, with Serial numbers H510946, H510651, H510190.

DateDec 19, 2006 (initiated Sep 21, 2006)
StatusTerminated
ReasonSelecting an empty space next to the ''Override'' or ''Acquire Actual'' buttons may result in mistreatment to the patient.
Root causeSoftware design
ActionOn September 21, 2006, the firm issued letters via certified mailing with return receipt to all its direct consignees, informing them of the affected product and providing instructions on the recall.
Quantity1689 units
DistributionNationwide. Defense Supply Center, VA or other Federal Government sales/distribution centers include: V.A. Medical Center, Atlanta, Dept. of Radiation Oncology, 1670 Clairmont Rd., Decatur, GA, 30030; V.A. Medical Center of Long Beach, Dept. of Rad. Oncology, 5901 E. 7th St., Long beach, CA 90745
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K001643VARIS 1.4G
Varian Medical Systems, Inc., Jun 23, 2000
K0504794D INTEGRATED TREATMENT CONSOLE
Varian Medical Systems, Mar 24, 2005