Recall Z-0153-2007 ยท BYJ Unit, Liquid-Oxygen, Portable
Precision Medical, Inc.
Precision Medical- Easy Mate Portable Liquid Oxygen System (container), PM22010 Series, Model number 2201, Precision Medical, Inc., Northampton, PA 18067
| Date | Nov 4, 2006 (initiated Oct 4, 2006) |
| Status | Terminated |
| Reason | Component assembly; The fill chuck is not assembled properly which may allow the device to come apart over time releasing its contents. |
| Root cause | Other |
| Action | The recalling firm telephoned the company to inform them of the problem and to make them aware that replacements would be made available starting on 10/16/06. |
| Quantity | 550 units |
| Distribution | The product was shipped to home care providers nationwide |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K041122 | PRECISION LIQUID OXYGEN SYSTEM Precision Medical, Inc., Oct 4, 2004 |