Recall Z-0115-2007 ยท JQP Calculator/Data Processing Module, For Clinical Use

Radiometer America Inc

Radiance Data Management System, Software Version 2.60, modification for the ABL800 Flex Blood Gas Analyzer

DateOct 28, 2006 (initiated Jul 6, 2006)
StatusTerminated
ReasonIncorrect FI02 values: programming issue can cause incorrect values to be transmitted to the LIS when : 1)- The FIO2 result is edited in the manual sample processing mode--2)- An existing result is opened and FI02 is then edited and sent.....In both cases RADIANCE will transmit the original FI02 value, not the value that was edited.
Root causeOther
ActionThe recalling firm initially contacted their customers and informed them of the software problem and the upcoming field correction for the software, via telephone on 7/6/2006. A follow-up letter dated 7/21/2006, was also sent to the customers.
Quantity5 units
DistributionWorldwide-The six USA units (equpped with Radiance software) were distributed to a total of two (2) U.S. customers: University of Missouri Hospitals and Clinics in Columbia, MO (4 units) and Vail Valley Medical Center, Vail, CO (2 units). World wide accounts pending submission. Three (3) Canadia
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K050869MODIFICATION TO ABL800 FLEX