Recall Z-0195-2007 ยท FDX Endoscopic Cytology Brush
Ekos Corp
Lysus Transport Stand (infusion stand), part number 4896-003
| Date | Nov 16, 2006 (initiated Oct 12, 2006) |
| Status | Terminated |
| Reason | Carts distributed without an additional caution label directly on the cart, alerting users to tipping hazard. |
| Root cause | Other |
| Action | The firm''s representatives will visit the consignees, apply the caution label and provide a letter, dated October 2, 2006, to the consignee. The letter states that the representative applied the caution label to the cart and reminds consignees if the UPS is removed and the cart is loaded with pumps and bags there is a potential for the cart to tip. |
| Quantity | 84 carts |
| Distribution | Nationwide. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.