Recall Z-0217-2007 ยท NJT Immunohistochemistry Reagents And Kits

Biogenex Laboratories

Super Sensitive Link-Label IHC Detection System/Mega Volume, RTU Multi-Link-HRP, REF/Cat. No. QP300-XAK, Kit containing reagent vials, Bio-Genex, San Ramon, CA.

DateDec 8, 2006 (initiated Oct 3, 2006)
StatusTerminated
ReasonMislabeling: The product's "contents section " labeling (outer box) is mislabeled incorrectly printed as '1 x 100 mL Alk Phos Label' which should read '1 x 100 mL HRP Label'. A second, smaller label, incorrectly reads "RTU MultiLink HPR/AEC" which should be labeled as "RTU MultiLink HPR". The reagent vials in the box are correctly labeled.
Root causeOther
ActionThe firms provided written notificaiton letter on approximately 10/03/06, which includes a ''reply back form'' which the firm has requested to be sent back to them. The firm has asked distributors to inform their accounts of the correction, in the event that the product has been shipped out to customers.
Quantity158 units
DistributionWorldwide-USA (including VA Medical Center) and Greece, Jordan, France, India, Taiwan, United Kingdom, Turkey, and Italy
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.