Recall Z-0186-2007 ยท FRN Pump, Infusion

Sorenson Medical, Inc.

ambIT Ambulatory Infusion Pump Cassette; Filter, Spike, Sterile, Vol. 2.3ml, REF 220140, Sorenson Medical, Inc., West Jordan, Utah.

DateNov 16, 2006 (initiated Oct 10, 2006)
StatusTerminated
ReasonSterile disposable cassette used with infusion pump may cause an under-infusion when used in a low flow rate therapy of less than 10 ml per hour.
Root causeOther
ActionConsignees were notified by letter on 10/10-11/2006 and told to immediately discontinue use/quarantine all affected cassettes. .
Quantity30,697 units
DistributionWorldwide, including USA, Belgium, France, Germany, Italy, Singapore, Venezuela.