Recall Z-0183-2007 ยท DQA Oximeter

Philips Medical Systems

M3001A Philips Multi Measurement Server (MMS)

DateNov 16, 2006 (initiated Sep 21, 2006)
StatusTerminated
ReasonUnexpected pulse oximetry (Sp02) readings (100%) over time when a sensor is not attached to the patient. In addition, with a sensor attached to a patient, an incorrect high Sp02 may be displayed when a pulse rate is in a range of 185BPM
Root causeOther
ActionPhilips issued a Customer Letter and Urgent Device Correction letter to consignees on 09/21/06. Users were notified of the problem and provided with Instructions to Mitigate Risk. A software upgrade to: version D.00.61 and Sp02 FW A.01.46 is mandatory.
Quantity69696 units
DistributionWorldwide, including USA, Canada, Australia, China, Hong Kong, India, Indonesia, Japan, Phillipines, Singapore, Taiwan, Thailand, Vietnam, Argentina, Brazil, Chile, Colombia, El Salvador, Jamaica, Mexico, Nicaragua. Panama, Paraguay, Peru, Venezuela, Austria, Belgium, Czech Republic, Denmark, Egypt,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K033715PICOSAT II SP02 PULSE OXIMETRY MODULE AND M3001A MULTI-MEASUREMENT SERVER
Philips Medical Systems, Inc., Feb 13, 2004