Recall Z-0183-2007 ยท DQA Oximeter
Philips Medical Systems
M3001A Philips Multi Measurement Server (MMS)
| Date | Nov 16, 2006 (initiated Sep 21, 2006) |
| Status | Terminated |
| Reason | Unexpected pulse oximetry (Sp02) readings (100%) over time when a sensor is not attached to the patient. In addition, with a sensor attached to a patient, an incorrect high Sp02 may be displayed when a pulse rate is in a range of 185BPM |
| Root cause | Other |
| Action | Philips issued a Customer Letter and Urgent Device Correction letter to consignees on 09/21/06. Users were notified of the problem and provided with Instructions to Mitigate Risk. A software upgrade to: version D.00.61 and Sp02 FW A.01.46 is mandatory. |
| Quantity | 69696 units |
| Distribution | Worldwide, including USA, Canada, Australia, China, Hong Kong, India, Indonesia, Japan, Phillipines, Singapore, Taiwan, Thailand, Vietnam, Argentina, Brazil, Chile, Colombia, El Salvador, Jamaica, Mexico, Nicaragua. Panama, Paraguay, Peru, Venezuela, Austria, Belgium, Czech Republic, Denmark, Egypt, |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K033715 | PICOSAT II SP02 PULSE OXIMETRY MODULE AND M3001A MULTI-MEASUREMENT SERVER Philips Medical Systems, Inc., Feb 13, 2004 |