ClearanceBrief
Recall Z-0837-2007 ยท MKJ Automated External Defibrillators (Non-Wearable)

Medtronic Emergency Response Systems, Inc.

LIFEPAK EXPRESS defibrillator

DateJun 21, 2007 (initiated Oct 5, 2006)
StatusTerminated
ReasonPremature battery depletion due to electrical component failure
Root causeNonconforming Material/Component
ActionPer request from Japan, the firm initiated action on 10/5/06 replacing the devices in Japan with new devices. On 3/2/07 the firm issued a Safety Alert letter dated February 2007 to consignees (with the exception of Japan), advising that the defibrillators have an identified, lead-free component that is susceptible to failure within the first 250 days after manufacture. Failure of this component might cause a short that will drain the battery and eventually render the device inoperative. The letter further states that the firm strongly recommends that the consignee immediately and on a reg
QuantityTotal of Express 1,006 units worldwide; amount does not include units distributed to Japan.
DistributionDistribution is worldwide. The device is designed for non-professional users and are primarily used in public access defibrillator environments such as airports, schools, and public libraries.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K011144LIFEPAK 600 AUTOMATED EXTERNAL DEFIBRILLATOR
Medtronic Physio-Control Corp., Dec 3, 2001