Recall Z-0049-2007 ยท IZL System, X-Ray, Mobile

Del Medical Systems Group

DynaRad Phantom Portable X-Ray System; Del Medical Imaging Corporation, 11550 West King Street, Franklin Park, IL 60131-1330; Phantom model (PH-150-CM & PH-150-G)

DateOct 19, 2006 (initiated Oct 3, 2006)
StatusTerminated
ReasonThe x-ray tube head assembly of the Phantom Portable X-Ray System may detach from the boom arm assembly. The tube head could fall and contact the patient or operator.
Root causeOther
ActionDel Medical sent Field Correction Recall letters dated 10/3/06 to the direct accounts, informing them of the potential for the shoulder bolt in the yoke assembly to come loose. If the shoulder bolt comes loose, the tube head could potentially fall and contact the patient or operator. The accounts were requested to locate the affected systems and take them out of service until they are inspected and, if needed, repaired.
Quantity53 units
DistributionNationwide and Thailand
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K941363PHANTOM
Dynarad Corp., Jul 20, 1994