Recall Z-0255-2007 ยท GBK
AGA Medical Corporation
Amplatzer Cardiac Septal Occluder Delivery Systems labeled as follows: a) AMPLATZER Delivery System, Order no. 9-DEL-5F-180/60, 9-DEL-6F-45/60, 9-DEL-6F-180/80, 9-DEL-7F-45/60, 9-DEL-7F-45/80, 9-DEL-7F-180/80, 9-DEL-8F-45/60, 9-DEL-8F-45/80, 9-DEL-8F-180/80, 9-DEL-9F-45/80, 9-DEL-9F-180/80, 9-DEL-10F-45/80, 9-DEL-12F-45/80. [Breakdown of order no. ''9-DEL-5F-180/60'' is as follows: 9-DEL, Fre
| Date | Dec 19, 2006 (initiated Oct 2, 2006) |
| Status | Terminated |
| Reason | AGA Medical has found that microscopic tears can occur in the delivery system sterile packaging under accelerated stress testing with routine shipping configurations. These microscopic tears are a potential breach of the sterile barrier. AGA has no confirmed complaints or adverse events related to this failure mode. |
| Root cause | Other |
| Action | An Urgent Recall Notice, dated 09/29/06, was sent to customers via facsimile or email and via Federal Express. The letter describes the issue and what product is affected. Unexpired product that is returned to AGA will be replaced. A response form is asked to be returned to AGA. |
| Quantity | 90,516 |
| Distribution | Worldwide, including USA, Puerto Rico, Algeria, Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Costa Rica, Cuba, Denmark, El Salvador, Finland, France, Germany, Greece, Guatemala, Hungary, Iceland, India, Indonesia, Israel, Italy, Japan, Jordan, Korea, Kuwait, Malays |
| FDA record | Open on fda.gov |