Recall Z-0172-2007 ยท KDI Dialyzer, High Permeability With Or Without Sealed Dialysate System
Asahi Medical Co., Ltd
Asahi Rexeed Series Multiple-Use Hollow Fiber Dialyzers; a wet model multiple use dialyzer consisting of hollow fiber membranes of polysulfone/PVP housed within a plastic casing of styrene copolymer, filled at the factory with fluid to facilitate priming by the user, and sterilized by gamma radiation before shipment; Asahi Medical Co., Ltd., 9-1, Kanda Mitoshirocho, Chiyoda-ku, Tokyo 101-8482, Jap
| Date | Nov 14, 2006 (initiated Oct 12, 2006) |
| Status | Terminated |
| Reason | There have been reports of difficulty separating the Gambro Cartridge Blood Set from Asahi Reuse/Multiple Use Dialyzers at the end of treatment, possibly damaging the dialyzer nozzle. |
| Root cause | Other |
| Action | Asahi sent Safety Alert letters on 10/12/06 to their re-use dialyzer customers advising them of the reports of difficulty separating the blood set from the dialyzer and not to use the Gambro Cartridge Blood Set with Asahi Reuse/Multiple Use Dialyzers in a reuse situation. Enclosed were copies of the temporary revision Instructions For Use with the following warning added at the top of the page: 'Additional Warning for Reuse/Multiple Use Dialyzers: Ashai Reuse/Multiple use dialyzers must not be used in connection with the Gambro cartridge Blood Set in a reuse environment.' Any questions were |
| Quantity | 100,000 for all models |
| Distribution | Nationwide |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K051187 | REXEED SERIES DIALYZERS Asahi Kasei Medical Co., Ltd., Jun 8, 2005 |