Recall Z-0172-2007 ยท KDI Dialyzer, High Permeability With Or Without Sealed Dialysate System

Asahi Medical Co., Ltd

Asahi Rexeed Series Multiple-Use Hollow Fiber Dialyzers; a wet model multiple use dialyzer consisting of hollow fiber membranes of polysulfone/PVP housed within a plastic casing of styrene copolymer, filled at the factory with fluid to facilitate priming by the user, and sterilized by gamma radiation before shipment; Asahi Medical Co., Ltd., 9-1, Kanda Mitoshirocho, Chiyoda-ku, Tokyo 101-8482, Jap

DateNov 14, 2006 (initiated Oct 12, 2006)
StatusTerminated
ReasonThere have been reports of difficulty separating the Gambro Cartridge Blood Set from Asahi Reuse/Multiple Use Dialyzers at the end of treatment, possibly damaging the dialyzer nozzle.
Root causeOther
ActionAsahi sent Safety Alert letters on 10/12/06 to their re-use dialyzer customers advising them of the reports of difficulty separating the blood set from the dialyzer and not to use the Gambro Cartridge Blood Set with Asahi Reuse/Multiple Use Dialyzers in a reuse situation. Enclosed were copies of the temporary revision Instructions For Use with the following warning added at the top of the page: 'Additional Warning for Reuse/Multiple Use Dialyzers: Ashai Reuse/Multiple use dialyzers must not be used in connection with the Gambro cartridge Blood Set in a reuse environment.' Any questions were
Quantity100,000 for all models
DistributionNationwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K051187REXEED SERIES DIALYZERS
Asahi Kasei Medical Co., Ltd., Jun 8, 2005