Recall Z-1485-06 ยท JAA System, X-Ray, Fluoroscopic, Image-Intensified

GE Healthcare

GE Precision RXi Digital Remote X-Ray Imaging R&F System. The Precision RXi digital is a remote R&F system consisting of a tilting table with integrated spot film device, 65 or 80 kW pulsing generator and high heat capacity X-Ray tube. The system can be configured to meet a wide variety of digital record and fluoroscopy exams for various clinical applications with either 32cm or 40cm image inten

DateSep 30, 2006 (initiated Nov 1, 2006)
StatusTerminated
ReasonAutomatic exposure control (AEC) automatically resets allowing another exposure without requiring manual reset as specified in 21 CFR 1020.31 (a)(3)(iv)
Root causeOther
ActionCorrection to devices is being implemented through GEHC Field Modification Instructions.
Quantity83
Distributionwithin US: CO, IL, LA, MA, MI, NC, PA, SC,TN, VA. OUS to include: Austrailia, Belgium, Burkina Faso, Brazil, Canada, China, Croatia, Czech Republic, Finland, France, Germany, Italy, Republic of Korea, Lebanon, Poland, Portugal, Russia, Slovakia, South Africa, Spain, Taiwan, Republic of China, Uni
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K041605PRECISION RXI ANALOG X-RAY SYSTEM
General Medical Merate S.P.A, Jun 30, 2004