Recall Z-0598-2007 ยท DTQ Console, Heart-Lung Machine, Cardiopulmonary Bypass

Terumo Cardiovascular Systems Corp

Terumo Advanced Perfusion System 1; 100V -120V, 15 A (circuit breaker), 50/60 Hz (20A power source required) base; Model 801763.

DateJun 30, 2007 (initiated Nov 1, 2005)
StatusTerminated
ReasonThe power manager subsystem may not initialize properly during start-up, which will cause the battery status monitoring and battery switching operations to not be fully functional.
Root causeOther
ActionFirm repaired the affected units beginning in November 2005, upon receipt of complaints of failure. The firm issued a Device Correction Letter dated 6/6/07 notifying all users of the problem and advising them that they will be visited and their units upgraded as replacement components become available.
Quantity485 of both model numbers
DistributionWorldwide distribution ---- Nationwide and countries of Australia, Belgium, Canada, Dominican Republic, Egypt, Germany, Hong Kong, India, Indonesia, Japan, Malaysia, Mexico, Pakistan, Philippines, Republic of Georgia, Russia and Saudi Arabia.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K022947ADVANCED PERFUSION SYSTEM 1 (100/120 VOLT),ADVANCED PERFUSION SYSTEM 1 (220/240 VOLT, ROLLER PUMPS (6, 4 INCH
Terumo Cardiovascular Systems Corp., Mar 20, 2003