Recall Z-0218-2007 ยท GWF Stimulator, Electrical, Evoked Response

Nihon Kohden America Inc

Nihon Kohden Electric Stimulator, Model number: MS-210BK (Optional accessories of MEB-2200A -Neuropack Evoked Potential and EMG Measuring System)

DateNov 28, 2006 (initiated Aug 23, 2006)
StatusTerminated
ReasonWhen the device is subject to physical shock, such as a drop to the floor, it may output a different value from the preset value. This may result in a superficial burn on the patient if the stimulation is done for a long time.
Root causeOther
ActionAll customers will be called by phone, then faxed the Product Notification starting 09/01/2006 Customers were instructed to indicate the serial numbers on each of your MS-210BK stimulators and fax the MS-210BK PRODUCT RECALLREPLACEMENT letter back to Nihon Kohden at: 949-580-15550 or email to customerservice@nkusa.com. Upon receipt of this letter, Nihon Kohden''s customer service department will send a replacement unit.
Quantity97 units in the U.S.
DistributionNationwide.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K991899NIHON KOHDEN MEB-2200A NEUROPACK EVOKED POTENTIAL & EMG MEASURING SYSTEM WITH ACCESSORIES
Nihon Kohden America, Inc., Sep 1, 1999