Recall Z-0314-2007 ยท DTQ Console, Heart-Lung Machine, Cardiopulmonary Bypass

Terumo Cardiovascular Systems Corp

Terumo Advanced Perfusion System 1; 100V -120V, 15 A (circuit breaker), 50/60 Hz (20A power source required); Models 801763.

DateDec 27, 2006 (initiated May 1, 2004)
StatusTerminated
ReasonThe gas system may fail calibration prior to use due to incorrect gas system flowmeter software.
Root causeProcess control
ActionU.S. consignees were visited and corrections made between May 2004 and November 2005. Foreign affiliates were instructed to make corrections on international units, but the firm has not received confirmation yet that these units have been corrected.
Quantity12
DistributionWorldwide distribution, including USA, Belgium, Germany, Japan, Korea and Republic of China.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K022947ADVANCED PERFUSION SYSTEM 1 (100/120 VOLT),ADVANCED PERFUSION SYSTEM 1 (220/240 VOLT, ROLLER PUMPS (6, 4 INCH
Terumo Cardiovascular Systems Corp., Mar 20, 2003