Recall Z-0306-2007 ยท DTQ Console, Heart-Lung Machine, Cardiopulmonary Bypass

Terumo Cardiovascular Systems Corp

Terumo Advanced Perfusion System 1; 100V -120V, 15 A (circuit breaker), 50/60 Hz (20A power source required); Model 801763.

DateDec 27, 2006 (initiated Nov 10, 2006)
StatusTerminated
ReasonImproperly adjusting the occlusion setting may result in a pump jam error and pump stoppage.
Root causeOther
ActionA letter dated 11/10/06 was sent to consignees providing them with an addendum (814576R/B) to the operators'' manual. Users of systems produced prior to May 2006 received only this addendum, but users produced between May and November 2006 had previously received addendum 814576R/A which is corrected and to be replaced by the new addendum 814576R/B.
Quantity488
DistributionWorldwide, including USA, Australia, Belgium, Canada, China, Dominican Republic, Egypt, Germany, India, Indonesia, Japan, Malaysia, Mexico, Pakistan, Republic of China, Singapore, South Africa, South Korea, Thailand and United Arab Emirates.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K022947ADVANCED PERFUSION SYSTEM 1 (100/120 VOLT),ADVANCED PERFUSION SYSTEM 1 (220/240 VOLT, ROLLER PUMPS (6, 4 INCH
Terumo Cardiovascular Systems Corp., Mar 20, 2003