Recall Z-0291-2007 ยท BSZ Gas-Machine, Anesthesia

Datex - Ohmeda, Inc

Datex Ohmeda Compact Absorber, Disposable, REF 427002100, white to violet and REF 427002000, Pink to white. Used with GE Healthcare ADU anesthesia systems. GE Healthcare Finland Oy, Helsinki, Finland +358 10 39411.

DateDec 19, 2006 (initiated Sep 11, 2006)
StatusTerminated
ReasonCertain Compact Absorbers may have an increased resistance to gas flow due to an improperly manufactured foam filter. The increased resistance can cause an elevated pressure at the ventilator end of the inspiratory circuit, but the pressure at the patient may be reduced. This could result in patient hypoventilation and hypoxia.
Root causeOther
ActionAll customers were contacted by letter in September 2006, and provided with a customer reply form. The firm sales some of the Compact Absorbers to distributors and they were asked to notify their Compact Absorber customers. Customers outside the US will be contacted to the end user and then their dealers and distributors. The letter said to immediately discontinue the use of all Compact Absorbers having a lot number as identified (reference codes 427002100 - white to violet color change) and (42700200 - pink to white color change). The affected lot numbers should be isolated and destroyed.
Quantity670,620
DistributionWorldwide, including USA, Austria, Australia, Belgium, Canada, Croatia, Germany, Denmark, Finland, France, Greece, Iceland, Ireland, Italy, Japan, South Korea, Kuwait, Netherlands, Norway, Poland, Portugal, Singapore, South Africa, Spain, Sweden, Switzerland, Turkey, United Arab Emirates, and Unit
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K973985AS/3 ANAESTHESIA DELIVERY UNIT (AS/3 ADU)
Datex-Engstrom, Inc., Oct 30, 1998