Recall Z-0034-2007 ยท JAK System, X-Ray, Tomography, Computed

Becton Dickinson & Company

BD Vacutainer Push Button Blood Collection Set with Pre-attached Holder. Non-pyrogenic 100. Rx only, sterile REF 368656 23G x 3/4'' x 12'' 0.6 x 19mm x 305mm ----- REF 367352 21G x 3/4'' x 12'' 0.8 x 19mm x 305mm

DateOct 12, 2006 (initiated Aug 11, 2006)
StatusTerminated
ReasonA complaint was received regarding a BD blood collection set where the blister was noticeably warped. The seal width was below minimum specification. Some blisters could adhere together and crack upon separation. Some units may have pre-activated or have defects affecting package integrity.
Root causeOther
ActionRecall notification were mailed out first class mail, return receipt requested, approximately 8/24/2006 to distributors, who were then asked to contact their customers.
Quantity110,300 units
DistributionNationwide.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K011984BD VACUTAINER PUSH BUTTON BLOOD COLECTION SET
Becton, Dickinson & CO, Aug 29, 2001