Recall Z-0159-2007 ยท DTQ Console, Heart-Lung Machine, Cardiopulmonary Bypass

MAQUET Inc.

Jostra HL-20 Heart Lung Machine - guiding pins

DateNov 7, 2006 (initiated Oct 3, 2006)
StatusTerminated
ReasonThe HL-20's pump head on the roller pump has four pairs of tube guide rolls that keep the tube in place in the raceway and roll along the inserted tube as the pump head turns. If dirt and other particles accumulate in the tube guide rolls, they cannot roll smoothly and finally they can get stuck.
Root causeOther
ActionAn 'Important Safety Notice - Please Read Immediately' letter was sent to all US customers on 10/3/06 by FedEx Return Receipt.
Quantity227 HL-20 base units with 545 single pumps
DistributionWorldwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K943803JOSTRA HL-20 INTEGRATED PERFUSION SYSTEM
Janus Biomedical, Inc., Jan 30, 1996