Recall Z-0158-2007 ยท DTQ Console, Heart-Lung Machine, Cardiopulmonary Bypass

MAQUET Inc.

Jostra - HL-20 Heart Lung Integrated Perfusion System, Roller Pump Hand crank

DateNov 7, 2006 (initiated Sep 1, 2006)
StatusTerminated
ReasonThe hand crank necessary for the emergency manual operation of the pump is too wide diametrically to fit into the holes which they were designed to fill.
Root causeOther
ActionThe firm sent a safety alert letter on 9/13/06 to all US customers by Federal Express Return Receipt.
Quantity132 or less in US. Firm officials stated that 66 base units in US 8/1/05 to 8/14/06 - each units comes with 2 hand crank
Distributionworldwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K943803JOSTRA HL-20 INTEGRATED PERFUSION SYSTEM