Recall Z-0005-2007 ยท HAW Neurological Stereotaxic Instrument
Stryker Instruments Div. of Stryker Corporation
Stryker Navigation System, eNlite System with Dell laptop computer model D800, Stryker # 7700-300-000.
| Date | Oct 3, 2006 (initiated Sep 8, 2006) |
| Status | Terminated |
| Reason | A component of the device is a Dell laptop computer, which contains batteries that are under recall because they can overheat and cause a fire. |
| Root cause | Other |
| Action | A recall letter dated 9/8/06 was sent to each customer instructing them to remove the battery from their Dell laptop computer model D800, to check the batteries to determine whether they are affected by the recall, and to only use the computer on AC power until they have requested and received replacement batteries. |
| Quantity | 218 units for Recalls Z-0005/0008-2007 |
| Distribution | Worldwide- USA and countries of Australia, Canada, France, Germany, Hong Kong, Italy, Japan, Mexico, New Zealand, South Korea, Switzerland and United Kingdom. |
| FDA record | Open on fda.gov |