Recall Z-1502-06 ยท LDD Dc-Defibrillator, Low-Energy, (Including Paddles)

Medtronic Emergency Response Systems, Inc.

LIFEPAK 20 external defibrillator/monitor

DateSep 13, 2006 (initiated Aug 30, 2006)
StatusTerminated
ReasonDevices with v38 system software do not display a ''LOW BATTERY: CONNECT TO AC POWER'' message when the monitor is on backup (DC) battery power and may shut down without warning.
Root causeOther
ActionOn August 30, 2006 the firm issued a letter dated August 2006 and titled MEDICAL DEVICE CORRECTION. The letter advises of the issue, explains how to resume power if power is shut down without warning, and advsies that a service representative will contact the consignee to make arrangements to install new software. On 10/19/06 the firm mailed an identical letter dated October 2006 to customers for units that were later identified as being subject to the recall.
Quantity2,831 worldwide
DistributionUnits were distributed worldwide to hospitals and medical centers.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K012274MEDTRONIC PHYSIO-CONTROL LIFEPAK 20 DEFIBRILLATOR/MONITOR/PACEMAKER
Medtronic Physio-Control Corp., Feb 5, 2002