Recall Z-1516-06 ยท JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use

Siemens Healthcare Diagnostics

ADVIA Centaur CP System, Automated Immunoassay Analyzer, Part /Catalogue Number 086-A001.

DateSep 23, 2006 (initiated Aug 4, 2006)
StatusTerminated
ReasonSample/Patient mis-identification (software defect)-the system can associate a test result and sample identification (SID) with an incorrect patient name when the Patient Demographics feature is used and the Patient Identification (PID) field is left blank
Root causeSoftware design
ActionOn 8/04/06, a Support Bulletin was e-mailed to all affected Bayer HealthCare LLC Branches worldwide in order to inform them of the issue and to provide them with instructions for this 'field correction' until the new software version is issued. A Customer Bulletin and Confirmation Form (FaxBack) were prepared and sent to the affected Bayer Healthcare LLC Branches on 8/04/06 for communication with the affected customers worldwide.
QuantityDomestic - 101, Foreign - 153
DistributionWorldwide