Recall Z-1304-06 ยท JAA System, X-Ray, Fluoroscopic, Image-Intensified

General Electric Med Systems LLC

GE Precision 500D Radiographic and Fluoroscopic Imaging System, Stationary X-ray System consisting of an X-ray generator; angulating table with x-ray Tube, collimator and image intensifier; wall stand; Overhead tube suspension; operator console and digital archive system.

DateSep 22, 2006 (initiated May 15, 2006)
StatusTerminated
ReasonSystems are non-compliant with Federal Performance Standard Title 21, Code of Federal Regulation (CFR), 1020.32(a)(1) in that the systems allow the production of x-rays when the primary protective barrier is not in position to intercept the entire cross section of the useful beam
Root causeOther
ActionAn Important Notice for Precision 500D, dated 11/24/2005 was hand delivered to all customers beginning 05/15/06, by GE field service engineers at the time they were to perform the correction to the device.
Quantity1,289
DistributionWorldwide, including USA, Puerto Rico, Canada, France, Germany, Indonesia, Italy, Japan, Jordan, Kuwait, Macedonia, Panama, Russia, Saudi Arabia, Singapore, and United Kingdom.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K011624EXPEDIO 500D R&F X-RAY SYSTEM
GE Medical Systems, Jun 8, 2001