Recall Z-1497-06 ยท FQO Table, Operating-Room, Ac-Powered
Skytron, Div. The KMW Group, Inc
Skytron General Purpose Surgical Table, battery model; Model 6600B, Dist. by Skytron, Grand Rapids, MI.
| Date | Sep 13, 2006 (initiated Aug 4, 2005) |
| Status | Terminated |
| Reason | The release levers can inadvertently release under load while the leg section is positioned at a ninety degree angle causing the leg section to drop. |
| Root cause | Other |
| Action | A service bulletin was issued to distributors on 8/12/05 instructing them to identify customers and to retrofit these tables. |
| Quantity | 160 total |
| Distribution | Nationwide. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K950192 | SKYTRON SKYBOOM SERIES Skytron, Div. the Kmw Group, Inc., Apr 3, 1995 |