Recall Z-0412-2007 ยท DSQ Ventricular (Assist) Bypass

Thoratec Corp

Thoratec Dual Drive Console a drive component of the Thoratec Ventricular Assist Device, Model Numbers, 10025-2600-005, 10025-2601-007, 10025-2602-006, Thoratec Corporation, 6035 Stoneridge Drive, Pleasanton, CA 94588

DateFeb 10, 2007 (initiated Sep 24, 2004)
StatusTerminated
ReasonThe module power switch (at the back of console) may break while attempting to turn the Dual Drive Console OFF or ON.
Root causeOther
ActionThe firm issued on 1/25/07 an URGENT: MEDICAL DEVICE CORRECTION-notice informing its consignees about the 'Power Switch Replacement Instructions' and arrangements for their service department replacement of the switches.
Quantity331
DistributionWorldwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.