Recall Z-1513-06 ยท FPO Stretcher, Wheeled
Stryker Medical Div. of Stryker Corporation
Stryker Power-PRO Powered Ambulance Cot, Model 6500.
| Date | Sep 22, 2006 (initiated Aug 29, 2006) |
| Status | Terminated |
| Reason | The head section wheel assembly may flex outward from the cot because the head section wheel bolts, which hold the assembly upright, may back out. |
| Root cause | Other |
| Action | Consignees were notified of the recall via letter dated 8/29/06. Included with the letter were replacement bolts for each cot and installation instructions, but the customer was given the option of contacting the recalling firm for installation. |
| Quantity | 1,841 cots (44 foreign) |
| Distribution | Worldwide distribution ---- including USA and countries of Canada, China, Germany and Switzerland. |