ClearanceBrief
Recall Z-1513-06 ยท FPO Stretcher, Wheeled

Stryker Medical Div. of Stryker Corporation

Stryker Power-PRO Powered Ambulance Cot, Model 6500.

DateSep 22, 2006 (initiated Aug 29, 2006)
StatusTerminated
ReasonThe head section wheel assembly may flex outward from the cot because the head section wheel bolts, which hold the assembly upright, may back out.
Root causeOther
ActionConsignees were notified of the recall via letter dated 8/29/06. Included with the letter were replacement bolts for each cot and installation instructions, but the customer was given the option of contacting the recalling firm for installation.
Quantity1,841 cots (44 foreign)
DistributionWorldwide distribution ---- including USA and countries of Canada, China, Germany and Switzerland.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.