Recall Z-0026-2007 ยท FNL Bed, Ac-Powered Adjustable Hospital

Extended Care Air Therapy Systems Inc

ECAT (Extended Care Air Therapy Systems) 2000 Safe Enclosure Bed

DateOct 14, 2006 (initiated Apr 5, 2006)
StatusTerminated
ReasonComponent defect- Following assembly and distribution, two (2) bolts may bend in the bed frame, which may allow the bed to tip into the wall
Root causeOther
ActionThe firm sent a Product Advisory Letter to their distributor/customers on 4/5/2006. The letter instructs the distributors as to the correct way to handle the beds during movement, loading, and unloading onto delivery trucks. The letter also urges the distributors to inspect the bed frame units in inventory at their locations for the presence of any bending, or damage to the four frame bolts. The letter states that the company will replace the bolts as soon as possible when ever necessary.
Quantity102 beds
DistributionThe recalled medical beds were distributed to six (6) wholesale/distributor customers located in the following states: KS, TN, MI, PA, and LA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.