Recall Z-0021-2007 ยท BSZ Gas-Machine, Anesthesia

Datex - Ohmeda, Inc

GE Healthcare-Datex-Ohmeda Aisys Anesthesia Delivery System with Aladin 2 cassettes, GE Healthcare, Madison, WI; Datex-Ohmeda, inc., a General Electric company

DateNov 7, 2006 (initiated Aug 10, 2006)
StatusTerminated
ReasonVaporizer shutdown with alarm-An intermittent leak in the Aladin 2 cassette/Aisys pneumatic interface (valves) may occur that will invoke an alarm state. When this occurs 'Vaporizer Failure' will appear on the Aisys main display, with an accompanying audible alrm. Delivery of anesthetic Agent is halted; gas flow and ventilation are not affected.
Root causeOther
ActionFirm mailed recall letters on August 10, 2006, to consignees telling them what to do if a failure occurred during system checkout or during use, as a short term solution. Later, GE representatives will make visits to each site for modification of the interface valve on each device, as the long term solution.
Quantity471
DistributionAK, AL, CA, FL, KS, KY, MD, MN, MO, NC, NE, SC, SD, TX, VA, WI,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K061609GE DATEX-OHMEDA AISYS
Datex-Ohmeda, Inc., Jun 27, 2006