Recall Z-1486-06 ยท IOR Wheelchair, Mechanical

Invacare Corporation

Kuschall''s K3/K4 Series of Manual Wheelchair, Model Airlite.

DateSep 6, 2006 (initiated Jul 18, 2006)
StatusTerminated
ReasonIf the user has chosen to install optional anti-tippers on the chair, the anti-tipper as designed may not be able to bear the stress of repeated or quick loads placed on it, causing the bolt to bend or break.
Root causeDevice Design
ActionThe recalling firm notified all consignees by recall letter sent on or about 7/20/06 along with replacement anti-tippers for each device on hand.
Quantity208 wheelchairs
DistributionNationwide and Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K982336KUSCHALL DESIGN K3/K4 SERIES OF MANUAL WHEELCHAIRS
Invacare Corp., Aug 21, 1998