Recall Z-1426-06 ยท DTZ Oxygenator, Cardiopulmonary Bypass

Cobe Cardiovascular, Inc

Cobe CML Duo Flat Sheet Membrane Oxygenator, sterile, (Custom Heart/Lung Packs), Catalog numbers: 067100001, 078002103, 078005802, 078159025, 078201013, 078212009, 078310006, 078332022, 078348014, 078428010, 078688007, 078869003. COBE Cardiovascular, Inc..

DateAug 24, 2006 (initiated Jun 26, 2006)
StatusTerminated
ReasonCooling and rewarming time is extended. Efficiency of the heat exchangers is reduced because many of the folds that are critical to heat exchange on the blood side of the units were blocked with a metalic material, thus decreasing the surface area needed for proper heat transfer.
Root causeOther
ActionConsignees were notified by letter on 06/26/2006.
Quantity580 units (Domestic)
DistributionWorldwide distribution --- USA including VA facilities in CO, GA, MO and countries of Australia, Canada, Italy, and Japan
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K000255COBE SMARXT OPTIMA XP SURFACE MODIFIED HOLLOW FIBER MEMBRANE OXYGENATOR
Cobe Cardiovascular, Inc., Apr 25, 2000
K963518COBE OPTIMA XP HOLLOW FIBER MEMBRANE OXYGENATOR
Cobe Cardiovascular, Inc., Dec 8, 1997