ClearanceBrief
Recall Z-1449-06 ยท JGS Electrode, Ion Specific, Sodium

Bayer Healthcare, LLC (Diagnostics Division)

ADVIA 1200 Chemistry System, Automated Clinical Chemistry Analyzer - human serum, plasma and urine tests.

DateAug 29, 2006 (initiated Feb 17, 2006)
StatusTerminated
ReasonLow QC recovery observed on carbamazepine (CARB) results immediately following a Gentamicin (GENT) assay or a Digoxin (DIG) assay. When GENT or DIG precedes CARB, the CARB test result is artificially low (as much as -25%). PHNY was determined to also exhibit a similar carryover effect on CARB. All others were found not to produce the same reagent probe-based carryover effect.
Root causeOther
ActionBayer HealthCare LLC issued a Support Bulletin to all affected Bayer HealthCare LLC Branches on/about 2/17/06 in order to inform them of the issue and to provide them with instructions for this ''field correction''. Customer Bulletins were also issued to the Branches both in the United States and elsewhere for communication with the affected customers.
Quantity1138 for all products
DistributionWorldwide including USA and Puerto Rico.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K990346BAYER ADVIA 1650 CHEMISTRY SYSTEM
Bayer Corp., May 21, 1999