ClearanceBrief
Recall Z-1446-06 ยท GKE Osmotic Fragility Testing

Becton Dickinson & Company

BD Unopette -- System Tests (Erythrocyte Fragility Test Kit). Catalog # 365830, Exp Date 2006-06, Manufactured by BD in San Lorenzo, Puerto Rico.

DateAug 26, 2006 (initiated Jun 1, 2005)
StatusTerminated
ReasonFormulation problems-lower pH of the solutions may increase Red Blood Cell hemolysis
Root causeOther
ActionRecall letters were sent via first class mail on 7/11/2005 by US mail, return receipt requested. A second notification was sent 8/4/2005
QuantityLot # 5014328 - 1,980 kits; Lot 5159075 - 4,360 kits
DistributionWorld wide-USA, Canada, Europe, Asia Pacific.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.