ClearanceBrief
Recall Z-1500-06 ยท HXJ Staple Driver

Xiros Plc

Neoligaments Staple Impactor, 6 mm Staple Impactor: 202-3001

DateSep 13, 2006 (initiated May 17, 2006)
StatusTerminated
ReasonThe impactors are being recalled for modification so that they can be disassembled fully for cleaning and sterilization, on suspicion that the assembled items retain contamination after cleaning and may not be effectively sterilized by the recommended hospital sterilization cycles.
Root causeOther
ActionCustomer recall notifications included: a) Telephone notification to highest volume users b) Recall letters were sent via certified mail to all customers June 1, 2006. Customers were instructed to cease using immediately and return the Impactors to Tapco. The customers were instructed to call the firm for a return authorization number. Modified loaner Impactors will be made available to customers as soon as possible.
Quantity89
DistributionNationwide & Worldwide, including Spain, Japan, UAE, Italy, Germany
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.