ClearanceBrief
Recall Z-1285-06 ยท HWT Template

Zimmer Inc.

Zimmer Trabecular Metal Shoulder Instrumentation Distal Pilot, 14 mm diameter, 170 mm length, non-sterile, catalog no. 4309-14-17 (00-4309-014-17).

DateJul 27, 2006 (initiated Jun 16, 2006)
StatusTerminated
ReasonDuring provisional trialing, the instrument may become stuck in the humeral canal and require surgical removal.
Root causeOther
ActionU.S. and international distributors were notified of the recall via letter dated 6/30/2006 to check customer inventories and to return recalled product to the firm. Trial implants were redesigned and reworked to eliminate the problem.
Quantity324 total
DistributionNationwide and worldwide including Australia, Canada, Germany, Italy and Spain.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.