Recall Z-1325-06 ยท CHL Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph

Radiometer America Inc

ABL800 Series Blood Gas Analyzer equipped with Software Version 5.21.

DateAug 3, 2006 (initiated Oct 31, 2005)
StatusTerminated
ReasonABL800 Series Blood Gas Analyzer became inoperable when the 'restore default setup' button was touched.
Root causeOther
ActionThe firm initiated this recall on 10/31/2005, at which time a Field Action Note was issued to their Service Reps. The notice instructed the service personnel to upgrade all ABL800 analyzer units presently running with software version V5.21 to the new version V5.23, in order to correct this problem. The Service Personnel began conducting visits at each customer site on 10/31/2005.
Quantity36
DistributionWorldwide, including USA and Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K041874ABL800 FLEX
Radiometer Medical Aps, Oct 8, 2004