Recall Z-1372-06 ยท NPQ Test, Factor V Leiden Mutations, Genomic Dna Pcr
Roche Diagnostics Corp.
Roche MagNA Pure LC System (RMD); Catalog no. 03670325001. An automated, general purpose instrument for isolation and purification of nucleic acids and assembly of PCR reactions.
| Date | Aug 5, 2006 (initiated Jul 17, 2006) |
| Status | Terminated |
| Reason | If the user opens the instrument lid during decontamination, the cycle will stop, but the timer will continue, resulting in the user believing the instrument has been decontaminated when it has not . |
| Root cause | Other |
| Action | Consignees were notified via recall letter dated 7/17/06 to restart the sterilization cycle if they open the instrument for any reason during the sterilization cycle. |
| Quantity | 771 total for recalls Z-1372-1374-06 |
| Distribution | Nationwide. |
| FDA record | Open on fda.gov |