ClearanceBrief
Recall Z-1372-06 ยท NPQ Test, Factor V Leiden Mutations, Genomic Dna Pcr

Roche Diagnostics Corp.

Roche MagNA Pure LC System (RMD); Catalog no. 03670325001. An automated, general purpose instrument for isolation and purification of nucleic acids and assembly of PCR reactions.

DateAug 5, 2006 (initiated Jul 17, 2006)
StatusTerminated
ReasonIf the user opens the instrument lid during decontamination, the cycle will stop, but the timer will continue, resulting in the user believing the instrument has been decontaminated when it has not .
Root causeOther
ActionConsignees were notified via recall letter dated 7/17/06 to restart the sterilization cycle if they open the instrument for any reason during the sterilization cycle.
Quantity771 total for recalls Z-1372-1374-06
DistributionNationwide.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.