Recall Z-1436-06 ยท KLK Monitor, Oxygen, Cutaneous, For Infant Not Under Gas Anesthesia

Radiometer America Inc

D826 Accessory kit for single tcpO2 Electrodes for use with the TCM400 Transcutaneous Monitor (Cutaneous Oxygen Monitor)

DateAug 31, 2006 (initiated Nov 1, 2005)
StatusTerminated
ReasonExcessive Drift-the membrane units of the device cause the electrode to exceed performance standards for drift in the first two or more calibrations
Root causeOther
ActionThe firm decided to initiate a recall on 9/8/2005 and the recall letter was issued in November, 2005. The letter informs the customers of this problem and asks that each customer inspect their current stock for the presence of the recalled lot codes and discard any such product found. The letter is accompanied by a recall response form which the customer can use to request additional replacement ellectrodes as necessary.
Quantity3,323-USA, 1,197-Foreign
DistributionWorldwide distribution --- including the states of PA, IL, TX, NY, FL, KS, NE, AL, MI, LA, WI, MN, NJ, CA, NH, MD, OH, MO, UT, IN, GA, WA, MA, KY, NC, IA, OK, SD, CO, AR, SC, ME, NV, AZ, RI, AK, MS, VA, WV, MT, TN, IN & NM. Country List pending
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K001866TCM400
Radiometer Medical A/S, Dec 7, 2000