ClearanceBrief
Recall Z-1323-06 ยท CHL Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph

Radiometer America Inc

Reference Electrodes for ABL 700/800 Blood Gas Analyzers. Electrode Part Number: 945-603, distributed one (1) electrode per box.

DateAug 3, 2006 (initiated Sep 19, 2005)
StatusTerminated
ReasonReference electrodes for the firm's ABL 700/800 Blood Gas Analyzers are oversized and will not fit into the instrument.
Root causeOther
ActionFollowing the receipt of three (3) customer complaints, on 9/14/2005, the firm issued an E-Mail notification [Management Cover] to their Service Personnel on 9/19/2005. The Management Cover instructed the service personnel to check the electrodes located at their stock and their customer locations; put all electrodes, which cannot be inserted in their call-back stock; and provide for credit to their customers through normal channels. The Service Personnel in turn, contacted their customers via telephone on or around 9/20/2005 and asked that they check their inventories for the presence of an
Quantity34
DistributionNationwide and Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K980130ABL700 SERIES BLOOD GAS , CO-OXIMETRY, ELECTROLYTE & METABOLITE SYSTEM
Radiometer America, Inc., Feb 9, 1998