ClearanceBrief
Recall Z-1384-06 ยท LFR Glucose Dehydrogenase, Glucose

Roche Diagnostics Corp.

Accu-Chek Inform blood glucose test strips; International Reference nos. 04535057001, 04647696001 and 04535120001.

DateAug 11, 2006 (initiated Jun 27, 2006)
StatusTerminated
ReasonThe drying agent beads, located in the cap, may become loose and fall into the vial causing the test strips to potentially give incorrect blood glucose test results
Root causeOther
ActionU.S. consignees, point of care coordinators or laboratory managers, long term care professionals, health care professionals, pharmacists, and mail order distributors were notified via letter dated 6/28/06 to cease using the strips and to contact the firm for replacement if they see loose beads in their vial. International accounts were also notified.
Quantity21,000,000 consumer vials and 10,000,000 hospital vials to U.S. plus 35,000,000 vials international-(Totals for Recalls
DistributionWorldwide Distribution-USA and countries of Argentina, Australia, Brazil, Canada, Chile, China, Colombia, Ecuador, England, France, Germany, Hong Kong, India, Indonesia, Japan, Malaysia, Mexico, New Zealand, Pakistan, Paraguay, Peru, Philippines, Singapore, South Africa, South Korea, Spain, Sweden,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.