Recall Z-1294-06 ยท LWP Implantable Pulse Generator, Pacemaker (Non-Crt)

Boston Scientific CRM Corp

Guidant INSIGNIA I Ultra --family of pacemakers includes the following: SR (models1190); and DR (models1290, 1291). Intermedics NEXUS I Ultra family of pacemakers includes the following: SR (model 1390) and DR (model 1490, 1491). The Intermedics NEXUS I Ultra (models 1390, 1490, 1491) are not available in the US. The INSIGNIA I Ultra pacemakers are multiprogrammable pacemakers from Guidant.--The

DateSep 28, 2006 (initiated Jun 24, 2006)
StatusTerminated
ReasonGuidant has determined that low-voltage capacitors from a single component supplier may malfunction in a manner that can lead to premature battery depletion or loss of pacing output without warning in the affected devices.
Root causeOther
ActionA Press Release was issued on 06/26/06. An Urgent Medical Device Safety Information & Corrective Action letter, dated 06/23/06 was delivered to physicians describing preliminary information before Guidant has completed its own investigation and finalized patient care recommendations. Non-implanted inventory is being retrieved from hospital shelves and Guidant Represenatatives. Physicians are to perform an in-clinic follow-up exam as soon as possible with patients implanted with affected devices.
Quantity18,406
DistributionWorldwide- USA and Antigua Barbuda, Argentina, Australia, Austria, Belgium, Canada, Chile, Colombia, Costa Rica, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Finland, France, Germany, Greece, Guadeloupe, Guam, Hong Kong, Iceland, India, Indonesia, Iran, Ireland, Israel, Ital
FDA recordOpen on fda.gov