Recall Z-1293-06 ยท LWP Implantable Pulse Generator, Pacemaker (Non-Crt)

Boston Scientific CRM Corp

Guidant INSIGNIA I Entra-- family of pacemakers includes the following: SSI (model numbers 0484 and 0485); DDD (models 0985, 0986); SR (models 1195, 1198); and DR (models1294, 1295, 1296). --Intermedics NEXUS Entra--family of pacemakers includes the following: SSI (model 1326); DDD (model 1426); SR (model 1398) and DR (model 1466, 1494). --The INSIGNIA I Entra pacemakers are multiprogrammable pace

DateSep 28, 2006 (initiated Jun 24, 2006)
StatusTerminated
ReasonGuidant has determined that low-voltage capacitors from a single component supplier may malfunction in a manner that can lead to premature battery depletion or loss of pacing output without warning in the affected devices.
Root causeOther
ActionA Press Release was issued on 06/26/06. An Urgent Medical Device Safety Information & Corrective Action letter, dated 06/23/06 was delivered to physicians describing preliminary information before Guidant has completed its own investigation and finalized patient care recommendations. Non-implanted inventory is being retrieved from hospital shelves and Guidant Represenatatives. Physicians are to perform an in-clinic follow-up exam as soon as possible with patients implanted with affected devices.
Quantity9,972
DistributionWorldwide- USA and Antigua Barbuda, Argentina, Australia, Austria, Belgium, Canada, Chile, Colombia, Costa Rica, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Finland, France, Germany, Greece, Guadeloupe, Guam, Hong Kong, Iceland, India, Indonesia, Iran, Ireland, Israel, Ital
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.