Recall Z-1440-06 ยท NDW Assay, Porphyrin, Spectrophotometry, Lithium

Beckman Coulter Inc

UniCel DxC 600/600PRO/600i/800/800PRO Synchron Clinical Systems , Part Numbers: A20463-Software version 1.0, A27331-Software version 1.2, A29764-Software version 1.4

DateAug 26, 2006 (initiated May 15, 2006)
StatusTerminated
ReasonIncorrect Reagent Status-When a new cartridge of Infinity Lithium Reagent is loaded on to the UniCel DxC System, the Reagent Status screen will incorrectly show 'Days Left' as 21. The correct Reagent Status for a new Lithium reagent cartridge is 14 days. Lithium reagent used past 14 days may produce low quality control and/or patient results.
Root causeOther
ActionA Product Correction Action (PCA) letter was sent the week of May 15, 2006 to all Unicel DxC customers informing them not to leave any Lithium reagent cartridge onboard a UniCel DxC System past 14 days. They also were instructed to complete and return the enclosed response form. The Letter was sent by US Mail.
Quantity524
DistributionNationwide and Canada
FDA recordOpen on fda.gov