Recall Z-1437-06 ยท KLK Monitor, Oxygen, Cutaneous, For Infant Not Under Gas Anesthesia

Radiometer America Inc

TC Monitor - Transcutaneous Monitor for p02 and pC02, Model #TCM4, (Cutaneous Oxygen Monitor)

DateAug 30, 2006 (initiated Feb 2, 2005)
StatusTerminated
ReasonAlarm may be silenced- Rapid switching, on and off, of the TCM4 monitor may cause the built-in acoustical alarm to remain silent when tripped.
Root causeOther
ActionThe firm initiated a recall action for the suspect device with a letter dated 2/2/2005. During a site visit to the U.S. customers conducted by the firm''s Field Service reps, the reps informed the user/customers of this problem and requested that all use of the device be discontinued, pending the formulation of a more permanent solution. Beginning on 2/8/2005, a second site visit was made to the same user/customers, at which time, the affected units were upgraded with operating software version 2.12. The firm''s recall/corrective action was scheduled for completion by 3/1/2005.
Quantity9 monitors were distributed.
DistributionNationwide