Recall Z-1261-06 ยท CHL Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph

Radiometer America Inc

Dri-Tek BG-OX Cartridge, Model #D7120, Cat. #946-003 Cartridges used in the NPT7 Analyzer, an in vitro diagnostic device.

DateJul 27, 2006 (initiated May 1, 2006)
StatusTerminated
ReasonAir bubbles may be trapped on or close to the O2 sensor in the pO2 chamber of the cuvettes in cartridges used for blood gas analysis on NPT7 analyzer. This produces too high pO2 readings without a question mark.
Root causeOther
ActionThe recalling firm initiated this recall with a letter to send to their customers, via Registered Mail, on 5/1/2006. The letter requests that the customers identify and discard the potentially defective cartridges from Lot #R0299. The letter is accompanied by a recall response page to be completed by the customer and faxed or mailed back to the Radiometer so that the customer can receive an equal number of replacement cartridges free-of-charge.
Quantity185
DistributionNationwide.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K982928NPT7 SERIES PH/BLOOD GAS AND CO-OXIMETRY SYSTEM
Radiometer America, Inc., Jan 6, 1999